NEW SQA and QwikCheck Cleaning Instructions | Effective May 2014

NEW SQA and QwikCheck Cleaning Instructions | Effective May 2014

The SQA cleaning procedure has been updated for ease of maintenance and for system reliability. It was found that improper or excessive cleaning with more than one drop of cleaning fluid may impact the system over time. Please note that the new procedure differs from the previous version in the following three ways:

• Use of the 6” wooden electrostatic brush “long brush” is now STEP 1 and should be done DAILY.

• Use of the “blue dot cleaning paddle” is different, is STEP 2 and is performed WEEKLY (instead of daily)

• New cleaning kits will include the updated instructions and will have an outside label indicating a new kit.

UPDATED CLEANING KIT INSERT (click to download):

Cleaning instructions PAGE 1 FINAL 5 27 14

BACK PAGE:

Cleaning instructions PAGE 2 FINAL 5 27 14