CAP Proficiency Testing on SQA Sperm Quality Analyzers

Prepare your team with MES guidance for SC-1 sperm count and PV-1 post-vasectomy proficiency testing, including sample preparation, capillary loading, and system-specific testing modes.

For tests not included in your CAP survey, contact MES for alternative assessment recommendations for Motility, Morphology, Vitality, and DFI testing.

Upcoming October Survey

Kit ships
October 19, 2026
Testing window
About 22 days
Exact deadline
In your CAP shipment

The October SC-1 Survey is scheduled to ship October 19, 2026. Laboratories are generally given 22 days to complete testing and report results.

Check the instructions and result forms included with your CAP shipment for your exact deadline. Review each mailing carefully, as directions may have changed since the previous survey.

Careful Preparation Supports Consistent Results

Proper sample handling, thorough mixing, and correct capillary loading are important steps before every test. Use the checklist below alongside the current CAP instructions and MES guidance for your system.

Review Steps, before you test...

1
2,800–3,200

Check REF. 2

Confirm the value is within the recommended range of 2,800–3,200.

2
15 min

Allow samples to reach room temperature

The MES sample-handling bulletin specifies 15 minutes.

3
30 sec

Mix thoroughly

Swirl and invert for 30 seconds. Do not vortex, and avoid creating bubbles.

4
Clean cup

Transfer carefully

Use a clean collection cup that accommodates the capillary, avoiding contact with the sides of the cup.

5
Mode + Level

Confirm your settings

Select the correct testing mode and applicable level for your system.

6
Every survey

Follow the current instructions

Review the CAP kit instructions before each survey.

Properly Load and Insert Your Capillary

Watch our video for a demonstration of proper capillary loading and insertion.

Before inserting the loaded capillary, confirm that:
  1. 1 The sample extends above the blue stopper into the cuvette.
  2. 2 The bottom of the black plunger is at the 0.5 mark.
  3. 3 No bubbles are visible, particularly in the back cube.

Wipe the capillary tip clean without drawing sample out of the tip, then insert the capillary into the testing chamber when prompted.

Select the Correct Testing Mode

SC-1 and PV-1 samples require different testing modes. Follow the instructions for your specific system.

SQA-Vision

SC-1 Sperm count
QC/Proficiency
PV-1 Post-vasectomy
Standard Post Vas
Including the manual testing option

Enter 1 mL for volume, follow the on-screen prompts, and complete the manual testing option. Do not report automated results alone.

SQA-iO

SC-1 Sperm count
QC/ProficiencyProficiency TestCAP/API
PV-1 Post-vasectomy
Standard Post Vas
Including the manual testing option

Enter 1 mL for volume, follow the on-screen prompts, and complete the manual testing option. Do not report automated results alone.

SQA-V

SC-1 Sperm count
QC / Stabilized Sperm

Follow the bulletin for V-Sperm setup and Level 1 selection. If testing reports 0.0, follow the bulletin’s confirmation procedure. Restore your routine QC settings after completing the survey.

PV-1 Post-vasectomy
Standard Post Vas
Including the Manual Capillary Scan

SQA-V Gold

SC-1 Sperm count
QC / Stabilized Sperm

Follow the bulletin for V-Sperm setup and Level 1 selection. If testing reports 0.0, follow the bulletin’s confirmation procedure. Restore your routine QC settings after completing the survey.

PV-1 Post-vasectomy
Standard Post Vas
Including the Manual Capillary Scan

Common Errors to Avoid

Selecting the wrong mode
Confirm the mode matches your system and survey sample.
Vortexing or insufficient mixing
Swirl and invert as directed, avoiding bubbles.
Incorrect capillary loading
Check sample level, plunger position, and the absence of bubbles.
Incomplete result review
Follow the manual testing and reporting instructions for your system.
Leaving survey settings in place
Restore routine QC settings after testing on SQA-V and SQA-V Gold systems.

Alternative Assessments Beyond Your CAP Survey

Does your laboratory perform Vitality or DFI testing? Contact MES for alternative assessment recommendations for tests not included in your CAP survey.

MES provides SQA-Vision recommendations covering assessment procedures, acceptance criteria, documentation, and Laboratory Director or qualified designee review.
When appropriate external proficiency testing is unavailable or does not evaluate your complete testing method, MES recommends documented alternative assessments at least semiannually, or at the frequency required by applicable regulatory and accreditation requirements.
Use formal external proficiency testing whenever an appropriate program is available and applicable to your method.

Downloadable Resources

Running CAP Samples on SQA Systems

System-specific instructions for sample preparation, testing modes, and reporting.

SQA-Vision Proficiency Testing & Alternative Assessment Recommendations

MES recommendations for external proficiency testing and semiannual assessments of Vitality and DFI.

Questions Before You Test?

Our team is here to help with system settings, sample preparation, capillary loading, and alternative assessment recommendations.