Technical Release Bulletin for QC of Morphology Pre-stained Slides
Background:
To perform quality control of the Morphology pre-stained smears please follow the instructions below.
Preparing QC semen samples:
- Pool around 5 FRESH semen samples and wash according to the following procedure:
- Mix the semen sample well.
 - Split into 5 equal volume samples.
 - Dilute each sample to 10 ml with 0.9% (9 g/l) saline or Earle’s buffer.
 - Centrifuge at 800g for 10 minutes.
 - Tip off and discard all but 20–40 μl of the supernatant.
 - Re-suspend the sperm pellet in the remaining supernatant by gentle pipetting.
 - Pool all washed samples.
 - Dilute the pool to 2 ml with 0.9% (9 g/l) saline or Earle’s buffer.
 - Add 20 μl of 35% formalin.
 
 - Distribute 50 μl aliquots of pooled semen samples into the screw cap cryo vials and store at 4 °C.
 - Assay sperm Morphology of these QC pool using the standard lab method to establish Morphology Target Range (Papanicolaou is recommended).
 - Establish pre-assayed Morphology Target Range: Mean Value +/- 3 standard deviations or 25% whatever is greater (please refer to the link below): 
http://www.mes-global.com/_blog/MES_Service_Blog/tag/Acceptable_Ranges_for_Semen_Analysis/ 
Morphology pre-stained slide QC procedure:
 • Assess the QC samples using Morphology pre-stained slides at pre-defined by the lab intervals according to the product insert procedure.
 • Compare Morphology QC results with the pre-assayed Morphology Target Range.
 • PASS/FAIL: Morphology QC results should fall into the pre-assayed Target Range.
Reference:
WHO laboratory manual for the examination and processing of human semen – 5th ed., World Health Organization 2010.
Compliance Date: Effective October 7th, 2017
Authority: Dr. Lev Rabinovitch, CTO
Distribution: Morphology pre-stained slide users
DOWNLOAD Technical Release Bulletin for QC of Morphology Pre-stained Slides
				